PRODUCTS

Extraction Oven System

No:OV-SCF-10000 Series

Info:

Application Focus
1. Chemistry / Chemical laboratory
2. Research institution
3. Chemical/ Food products factories
4. Trace of extract sample

• Product Description •

Oven Extraction Furnace
Specifications
1. Operation temperature: room temperature ~220 ℃
2. Operation pressure: Operating pressure up to 7250 psi (MWAP=7250 PSI)
3. Operation capacity: Optional 60 / 250 / 1000ml High-pressure extraction vessels
4. Double extraction loop: Optional 1 or 2 extraction loop
5. Extraction mode: dynamic and soak extraction
6. Cleaning method: capable of solvent injection device
7. Flow control: Adjustable flow rate control with display of CO2 flow rate
8. Certification: Whole system pass CE
        IEC 61010-2-010 : 2002 (2nd Ed) for use with IEC 61010-1 : 2001
         EN 61010-2-010 : 2004 (2nd Ed) for use with EN 61010-1 : 2002 Standard
9. Dimension: 700 mm × 650 mm × 950 mm
10.Weight: 80 kg

Utility Needs
1. Power supply: 110V , 16A

Application Focus
1. Supercritical , liquid CO2 extraction
2. Provide column chromatography or separation column for hot air heating

---------------------------------------------------------------------------------------------------------------
Digital Pressure Control Module(PM-10000C)
Specifications
1. System output pressure: 1500~10000 psi
    (depending on driving pressure)
2. System output flow rate: 200ml/min (up to 4350 psi)
3. Booster pump
4. Built-in pre-cooler (without chiller)
5. Piping: seamless stainless steel tube
6. Certification: Whole system pass CE
    IEC61010-2-010(2nd Ed) for use with IEC61010-1:2001  Standard
7. Dimensions: 360 mm × 600 mm × 270 mm
8. Weight: 30 kg

Control System
1.Touch-Panel for Pressure Control Systems and human-machine operation interface
2. Programmable pressure control system
3. Proportional precise pressure control
4. Multi-step for booster and immediately adjust / response control
5. Alarm (notice) for abnormal operation and operation timer

Safety Protection
1. Overpressure abnormal alarm devices
2. Emergency stop devices and automatic gas cut-off installations for overpressure
3. Safety relief valve for overpressure (full flow rate)

Utility Needs
1. Power supply: 110/220V
2. Source air: (5~8 kg/c㎡)
3. Chilly: water temperature≤5℃

Application Focus
1. Liquid CO2 transmission and booster
2. Supercritical CO2 cleaning equipment applications
 

Contact Taiwan Supercritical Technology Co., Ltd.

Contact Us Now